Pagina 1 - PORTABLE NEBULIZER

INSTRUCTION MANUALRPORTABLE NEBULIZERMODEL NE-U03V

Pagina 2 - INTRODUCTION

10CLEANINGBe sure unit is OFF and the power cord is disconnected before cleaning.Nebulizer Parts and Casing(EXCEPT MESH CAP)There are several ways to

Pagina 3 - IMPORTANT SAFEGUARDS

11MESH CLEANINGNOTE: The mesh cap is the most delicate part of the MicroAir. It is importantnot to directly touch the mesh cap. Follow the cleaning in

Pagina 4 - KNOW YOUR UNIT

12STORAGE1. Completely dry all cleaned and sterilized parts.104°F50°FMesh CapAttach the mesh cap to the unit as describe inthe operating instructions.

Pagina 5 - CAUTIONS - WARNINGS

13Using BatteriesBATTERY OPERATION1. Turn the unit upside down. The battery lid willlift off by pressing down towards the arrows.2. Insert four (4) &

Pagina 6

Symptom Check Remedy14* Do not try to repair unit - no parts of the unit are user serviceable. TROUBLE SHOOTING GUIDE(check before sending for service

Pagina 7 - UNIT ASSEMBLY

15Medication Cups (MODEL #U03-7)Aerosol Mask (MODEL #U03-2)A 5 ml Medication Cup Plastic Aerosol MaskMouthpiece (MODEL #U03-3)Mesh Cap (MODEL #U03-4)P

Pagina 8 - OPERATING INSTRUCTIONS

16SPECIFICATIONS*Power Source: AC 120v 60 HzPower Consumption: Approximately 1.2 wVibration Frequency: Approximately 65 KHzNebulization Rate: 0.25 ml/

Pagina 9

17COMMON QUESTIONS ABOUT THE MICROAIRQ. Can I use any type of double “AA” batteries in theMicroAir?A. We recommend that alkaline batteries be used for

Pagina 10 - CLEANING

COMMON QUESTIONS ABOUT THE MICROAIRQ. Will the mesh break easily?A. The mesh is like all small delicate parts that can be broken if not gentlyhandled.

Pagina 11 - B. Daily Cleaning

19TWO YEAR LIMITED WARRANTYYour Omron MicroAir Portable Nebulizer is warranted to the original retailpurchaser to be free from defects in material and

Pagina 12

TABLE OF CONTENTSIntroduction………………………………………………………………………2Important Safeguards ……………………………………………………………3Know Your Unit …………………………………………………………………4Caution

Pagina 13 - BATTERY OPERATION

RDistributed byOMRON HEALTHCARE, INC.300 Lakeview ParkwayVernon Hills, Illinois 60061Phone: 1-800-634-4350www.omronhealthcare.comCopyright © OMRON HEA

Pagina 14 - TROUBLE SHOOTING GUIDE

IMPORTANT SAFEGUARDSWhen using electrical products, especially when children are present, basicsafety precautions should always be followed, including

Pagina 15 - AVAILABLE ACCESSORIES

4KNOW YOUR UNIT* The shape of the plug may vary from country to country.AC Adapter*UnitCoverMaskMedicationCupMesh Cap(Extra Cap Included)MicroAir Unit

Pagina 16 - SPECIFICATIONS*

5CAUTIONS - WARNINGS1. WARNING The most delicate part of the unit is the mesh cap.NEVERtouch the mesh cap directly. Follow the cleaning instructionsex

Pagina 17

6CAUTIONS - WARNINGS10. WARNING Fill the medicationcup before usingthe unit. Do notshake the unit dur-ing use.9. WARNING Do not submerge theunit in w

Pagina 18

7How to AssembleA. Clean and sterilize all unit parts before the first use or after a long storage period.B. Fill the medication cup before plugging i

Pagina 19 - TWO YEAR LIMITED WARRANTY

8OPERATING INSTRUCTIONS1. Before Operating Make sure all parts are mounted properly.Check to fill the medication up to correct level of 5 ml.2. Plug

Pagina 20 - OMRON HEALTHCARE, INC

9NOTE: The mesh cap may be damaged by hardened medication that is left inthe vibration horn effecting how the unit nebulizes. The following are mainte