Pagina 1 - INSTRUCTION MANUAL

Automatic Blood Pressure Monitorwith ComFit™CuffModel HEM-780INSTRUCTION MANUAL780®ENGLISH ESPAÑOL Final ES HEM-780 IM 1/28/05 1:11 PM Page 1

Pagina 2 - CONTENTS

The Blood Pressure Monitor automatically stores up to 90 measurementvalues with the date and time. The monitor also calculates an averagereading based

Pagina 3 - SAVE THESE INSTRUCTIONS

3. SETTING THE MONTHPress the memory up button to advance by increments of one month.Press the Date/Time Setting button to set the current month.The

Pagina 4 - IMPORTANT SAFETY NOTES

6. SETTING THE MINUTEPress the memory up button to advance by increments of one minute.Press the Date/Time Setting button to set the current minute.T

Pagina 5 - BEFORE TAKING A MEASUREMENT

1. Make sure the air plug is securely inserted in the main unit.2. Remove tight-fitting clothing from your upper arm.3. Sit in a chair with your feet

Pagina 6 - KNOW YOUR UNIT

144. Hold the grip on the cuff securelywith your hand.5. Turn the palm of your hand upward.6. Apply the cuff to your left upper arm so the blue strip

Pagina 7 - UNIT DISPLAY

157. Wrap the cuff firmly in place around your arm using the velcro strip.APPLYING THE CUFF ON THE RIGHT ARMWhen taking a measurement using the right

Pagina 8 - BATTERY INSTALLATION

The monitor automatically determines your ideal inflation level. In rarecircumstances when a higher inflation may be necessary, the monitorautomatical

Pagina 9 - USING THE AC ADAPTER

You3blood pressure and pulse rate are dis-played.4. Press the START/STOP button to turnthe monitor off.NOTE: The monitor will automaticallyturn off af

Pagina 10 - SETTING THE DATE AND TIME

Every time you complete a measurement the monitor stores the bloodpressure and pulse rate in the memory. The Blood Pressure Monitorautomatically store

Pagina 11

TO DISPLAY THE MEASUREMENT VALUESThe date and time is alternately displayed with the measurement values.1. Press the memory down button to display th

Pagina 12

Before Using the MonitorIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3Important Safety Notes . . .

Pagina 13 - APPLYING THE ARM CUFF

To keep your digital blood pressure monitor in the best condition andprotect the unit from damage follow the directions listed below:Keep the monitor

Pagina 14

SYMBOL CAUSE CORRECTIONMonitor did not detectpulse rateRemove the arm cuff. Refer to“Applying the arm cuff” on page13. Wait 2-3 minutes. Take another

Pagina 15

22TROUBLESHOOTING TIPSPROBLEM CAUSES AND SOLUTIONSNo power.No display appears on the unit.Replace worn batteries.Check the battery installation for pr

Pagina 16 - TAKING A MEASUREMENT

23QUICK REFERENCE GUIDETo ensure a reliable reading, avoid eating, smoking, and exercising for30 minutes before taking a measurement. You should try t

Pagina 17

245. Press the START/STOP button.Display symbols appear. The cuff starts toinflate automatically.Remain still and do not talk during the measurement.N

Pagina 18 - USING THE MEMORY FUNCTION

25FCC STATEMENTNote:POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)This product has been tested and found to comply with the limits for

Pagina 19

WARRANTY INFORMATIONYour HEM-780, IntelliSense Automatic Blood Pressure Monitor, excluding themonitor cuff, is warranted to be free from defects in ma

Pagina 20 - CARE AND MAINTENANCE

27SPECIFICATIONSModel: HEM-780Display: LCD Digital DisplayMeasurement Range: Pressure: 0 to 299 mmHg, Pulse: 40-180/min.Accuracy/Calibration: Pressure

Pagina 21 - ERROR INDICATORS

Manufactured by:OMRON HEALTHCARE, INC.1200 Lakeside DriveBannockburn, Illinois 60015www.omronhealthcare.comMade in ChinaCopyright © 2004 Omron Healthc

Pagina 22 - TROUBLESHOOTING TIPS

Thank you for purchasing Omron’s HEM-780 IntelliSense®AutomaticBlood Pressure Monitor with the ComFit™Cuff.INTRODUCTION3Fill in for future reference.D

Pagina 23 - QUICK REFERENCE GUIDE

To assure the correct use of the product basic safety measures shouldalways be followed including the precautions listed below: þ Read all informatio

Pagina 24

5BEFORE TAKING A MEASUREMENTTo ensure a reliable reading follow these recommendations:1. Avoid eating, smoking, and exercising for 30 minutes before t

Pagina 25 - FCC STATEMENT

6KNOW YOUR UNITAir JackMain Unit:Battery CompartmentStart/StopButtonMemory ButtonsDate/Time Setting ButtonDisplayAC Adapter JackComponents:Arm CuffAir

Pagina 26 - WARRANTY INFORMATION

7UNIT DISPLAYDiastolic Blood PressureSystolic Blood PressureMemory SymbolPulse DisplayDate/Time DisplayBattery Low SymbolAverage Value SymbolHeartbeat

Pagina 27 - SPECIFICATIONS

81. Press the ▼ indicator on the battery coverand slide the cover off in the direction ofthe arrow.2. Install 4 “AA” size batteries so the + (positive

Pagina 28 - HEM-780INST Rev A

Use only the authorized Omron AC Adapter, Model Number HEM-ADPT1,with this monitor. Use of any other adapter may cause damage to themonitor and will v